Oklahoma Administrative Code (Last Updated: March 11, 2021) |
TITLE 252. Department of Environmental Quality |
Chapter 410. Radiation Management |
Subchapter 11. Use of X-Rays and /or Particles for Therapeutic Purposes in the Healing Arts and Veterinary Medicine |
Part 3. THERAPEUTIC SYSTEMS OF LESS THAN 1 MEV |
SECTION 252:410-11-36. Control panel functions
Latest version.
- (1) an indication of whether electrical power is available at the control panel and if activation of the x-ray tube is possible;(2) an indication of whether x-rays are being produced;(3) means for indicating x-ray tube potential and current;(4) means for terminating an exposure at any time;(5) a locking device which prevents unauthorized use of the x-ray system; and(6) for post-1989 systems, a positive display of specific filters in the beam.