SECTION 252:410-11-36. Control panel functions  


Latest version.
  •   The control panel of a <1 MeV system must have:
    (1)   an indication of whether electrical power is available at the control panel and if activation of the x-ray tube is possible;
    (2)   an indication of whether x-rays are being produced;
    (3)   means for indicating x-ray tube potential and current;
    (4)   means for terminating an exposure at any time;
    (5)   a locking device which prevents unauthorized use of the x-ray system; and
    (6)   for post-1989 systems, a positive display of specific filters in the beam.
[Source: Added at 17 Ok Reg 1136, eff 6-1-00]