Oklahoma Administrative Code (Last Updated: March 11, 2021) |
TITLE 252. Department of Environmental Quality |
Chapter 410. Radiation Management |
Subchapter 11. Use of X-Rays and /or Particles for Therapeutic Purposes in the Healing Arts and Veterinary Medicine |
Part 3. THERAPEUTIC SYSTEMS OF LESS THAN 1 MEV |
SECTION 252:410-11-37. Multiple tubes
Latest version.
- When a control panel of a <1 MeV system can be used to energize multiple tubes, the operator must be able to activate only one x-ray tube at a time and confirm which x-ray tube is activated by viewing a control panel display. The machine must also have a display on each tube housing assembly which confirms that a specific tube is energized.