SECTION 252:410-11-73. Calibrations of <1 MeV systems  


Latest version.
  • (a)   Frequency. The calibration of a <1 MeV system shall be performed at intervals not to exceed one (1) year and after any change or replacement of components which could cause a change in the radiation output.
    (b)   Dosimetry system. The calibration of the radiation output of <1 MeV systems shall be performed with a dosimetry system that has been calibrated within the preceding two (2) years with calibration that is traceable to a national standard.
    (c)   Performance. The therapy system calibration shall be performed by or under the direction of a qualified expert.
    (d)   Precision. Calibrations must be in sufficient detail so that the dose at a reference point in water or plastic phantom can be calculated to within an uncertainty of 5 percent.
    (e)   Outcome. Calibrations must include, but are not be limited to, the following determinations:
    (1)   verification that the system is operating in compliance with the design specifications;
    (2)   half-value layer for each kV setting and filter combination used;
    (3)   exposure rates as a function of field size, technique factors, filter, and treatment distance used; and
    (4)   the degree of congruence between the radiation field and the field indicated by the localizing device if such device is present which shall be within 5 millimeters for any field edge.
[Source: Added at 17 Ok Reg 1136, eff 6-1-00]