Oklahoma Administrative Code (Last Updated: March 11, 2021) |
TITLE 310. Oklahoma State Department of Health |
Chapter 281. Diagnostic X-Ray Systems |
Subchapter 11. Use of X-Rays in the Healing Arts and Veterinary Medicine |
SECTION 310:281-11-11. Additional operational controls for veterinary x-ray systems
Latest version.
- (b) Unless required to restrain an animal, the operator shall stand at least 6 feet away from the useful beam and the animal during radiographic exposures.(c) No individual, other than the operator, shall be in the x-ray room or area while exposures are being made unless such individual's assistance is required.(d) When an animal must be held in position during radiography, mechanical supporting or restraining devices shall be used when technique permits.(1) If the animal must be held by an individual, that individual shall be protected with appropriate shielding devices, such as protective gloves and apron, and the person shall be so positioned that no part of his/her body except hands and arms will be struck by the useful beam.(2) The exposure of any individual who holds animals shall be monitored by film badge, TLD, or other appropriate device.